Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-67
Figure 8-53. Power assembly pulled out
8.14.3 Line filter
Remove the line filter as follows. Install by reversing removal procedure, referring to
Figure 9-24 as necessary.
1.
Remove power assembly (Section 8.14.2).
2.
Disconnect wires from line filter (Figure 8-54).
3.
Remove two M3 x 8 flat-head, POZIDRIV™* screws that secure line filter. Remove line
filter.
NOTE:
When installing line filter, make sure ground terminal is at bottom.
Figure 8-54. Line filter, circuit breaker, and power switch connections
7-00128
Circuit breaker
Power cord
retainer
Power cord
retainer screw
Line filter
7-00213
Circuit
breaker
Power
switch
Line filter