Self-tests
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
3-9
NOTE:
When EST requires a user response, the ventilator will wait indefinitely.
5.
The ventilator automatically starts the first test in the sequence (Table 3-8). The tests all
run automatically and consecutively, unless a fault occurs. Some tests display additional
prompts requiring your response. EST prompts (informational prompts and those
requiring your response) are listed in Table 3-6.
At the end of each test, the test’s name and pass/fail/fault status are displayed. If an EST
test does not pass, you may still continue EST execution, although the entire EST will not
pass until the error conditions are corrected.
POST running...
Wait a few seconds until POST is completed.
NOTE:
• If a single beep is not audible during POST, the main audible alarm may be
malfunctioning. Troubleshoot as for EST main alarm test (Table 3-9).
• It is best not to power off the ventilator during POST.
humidification device type
Choose humidifier
Turn knob to select desired humidification device, then press ACCEPT. Device types
include:
HME
(heat-and-moisture exchanger or “artificial nose”),
Dual heated wire
(humidifier with heated wires on both inspiratory and expiratory limbs), or
No
heated wire
(humidifier without a heated wire on expiratory limb).
Warning
Incorrectly specifying the humidifier type during EST can affect the accuracy
of spirometry calculations.
tubing type
Choose tubing type
Turn knob to select either
Adult tubing
or
Pediatric tubing
and then press ACCEPT.
Warning
Incorrectly specifying the ventilator breathing circuit type during EST can
cause an inappropriate sensitivity for the leak test and occlusion alarm.
Covidien recommends using pediatric circuits when ventilating patients with
5 mm or smaller internal-diameter artificial airways.
NOTE:
To ensure compliance compensation functions correctly, the user must run
EST with the circuit configured as intended for use on the patient.
ET size: x.x mm
Choose ET size
Turn knob to select appropriate ET (endotracheal tube) size, and then press
ACCEPT.
Warning
Specifying an ET tube size too large for small pediatric patients can cause
premature termination of breaths.