Diagnostic codes
6-10
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
6039
POST (technical alert --
VALVES TEST FAILED)
Safety valve stuck closed
during POST
1. Check state of safety valve during POST.
If stuck closed, replace.
2. If safety valve open during POST,
replace cylinder outlet check valve.
6040
Ongoing hardware
checks
Nebulizer failed alert, power
not being supplied to
nebulizer.
1. Check cable connections between the
communications panel and the
nebulizer.
2. Replace the cable.
3. Check cable connections between the
communications PCB and the controller
PCB.
4. Replace the cable.
5. Replace communications PCB.
6. Replace the nebulizer.
6041
BIOT (technical alert --
PARTIAL OCCLUSION)
Partial occlusion in one limb of
the breathing circuit.
1. Check patient.
2. Check limbs for blockages.
3. Check exhaust port for blockage.
4. Replace pressure solenoid PCB.
6042
BIOT (technical alert --
PCYL COMPARE
ALLERT)
P
cyl
transducer tube is
incorrectly connected to P
i
or
P
e
transducer.
1. Check tubing connections.
2. Replace pressure solenoid PCB.
6043-6046
BIOT (technical alert --
CONTACT SERVICE)
BBU signal conflict: signals
indicate both battery and AC
power operation
1. Verify AC power to the ventilator.
2. Verify battery voltage.
3. Check that connections between the
BBU and controller PCBs are secure.
4. Replace BBU PCB.
5. Replace controller PCB.
7000
POST
Software error
7001
POST
Software error
7002
POST
Software error
7003
POST
Microcontroller or other
hardware failure
Replace controller PCB.
7004
POST
Microcontroller or other
hardware failure
Replace controller PCB.
7005
POST
Microcontroller or other
hardware failure
Replace controller PCB.
7006
POST
Microcontroller or other
hardware failure
Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action