Diagnostic codes
6-4
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
6.4.3 Error code 09169
Revision K software for the 700 Series Ventilators was
released in October, 1999. Error code
09169 only affects units with Revision J Software.
The ventilator is designed to deliver RAMP and SQUARE flow waveform VCV (Volume
Controlled Ventilation) in A/C (Assist Control) and in SIMV (Synchronous Intermittent
Mandatory Ventilation). The flow waveform is user selectable via the Menu Key under
Option 4, User Settings. When the ventilator is in SIMV mode with Ramp VCV breath type
selected with set tidal volumes greater than 200 ml, the ventilator may deliver tidal volumes
outside the stated volume delivery accuracy.
This anomaly for RAMP VCV has been corrected in Revision “K” Software.
J Level Software currently installed in any 700 Series Ventilator, is to be removed.
Table 6-1: Diagnostic codes
Code
Software
Possible cause
Corrective action
1000-1177
UI application
Software error
1178
UI application
The unit was shut down after
low power was detected. The
AC supply was lost and all
batteries in unit are exhausted.
1. Restore AC power, if possible.
2. Replace batteries.
1179-1xxx
UI application
Software error
2000-2xxx
UI driver
Software error
3000-3xxx
BDMC application
Software error
4000-4xxx
BDMC driver
Software error
5000-5012
Interprocessor
communication
Software error
5013
Interprocessor
communication
Microcontroller or other
hardware failure
Replace controller PCB.
5014
Interprocessor
communication
Software error
Microcontroller failure
See Section 6.2.
Replace controller PCB.
5015
Interprocessor
communication
Microcontroller or other
hardware failure
Replace controller PCB.
5016-5xxx
Interprocessor
communication
Software error
6001
Built in ongoing tests
(BIOT) (technical
alert -- LOSS OF
POWER)
The AC supply was lost and the
batteries are so low that all
power may be lost. The unit
will shut down in 5 minutes.
1. Restore AC power, if possible.
2. Check battery connections.
3. Replace batteries (internal and
external, if installed) as necessary.
4. Replace BBU PCB.
NOTE:
Actual battery backup time remaining depends on battery condition
and ventilator settings.