Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-39
12004
Communications
routines
Corrupt data packet. Five
unsuccessful attempts were
made to transmit it.
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12005
Communications
routines
Software error
12006
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12007
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12008
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12009
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12010
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12011
Communications
routines
Communications failure
1. Verify communications configuration
on PC side.
2. Check communications cabling.
12012
Communications
routines
Software error
12013
Communications
routines
Communications failure
Replace controller PCB.
12014
Communications
routines
Software error
13000
Ventilator status
Ventilator reset occurred while
entering SST
13001
Ventilator status
Ventilator reset occurred while
entering EST
13002
Ventilator status
Ventilator reset occurred
during EST
13003
Ventilator status
Ventilator reset occurred
during SST
13004
Ventilator status
Ventilator reset occurred
during calibration
13005
Ventilator status
Ventilator reset occurred in
standby mode
13006
Ventilator status
Ventilator reset occurred in
normal ventilation mode
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action