Performance verification
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
5-3
5.3 Preliminary ventilator cleaning and inspection
Warning
To prevent disease transmission, use personal protective equipment when handling
contaminated bacterial filters or other patient accessories. Refer to the
700 Series
Ventilator System Operator’s Manual
for instructions on sterilizing patient system
parts.
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
servicing components inside the ventilator.
NOTE:
If you find any problems during preliminary ventilator inspection, correct them before
proceeding with performance verification. Failure to correct such problems now may
affect the remainder of performance verification.
Ventilator breathing circuit
To use as test circuit: G-061208-00 or
equivalent (adult, reusable, without
heated wire)
NOTE:
To ensure compliance compensation
functions correctly, you must run EST
or SST with the circuit configured as
intended for use on the patient.
Test lung with strap
4-000612-00
Tubing kit, performance verification (Includes
coupling (P/N 4-003443-00), tube junction connector
(P/N 4-011521-00) (quantity of 2), and 30-cm, 3/16-in.
ID x 3/8-in. OD tubing)
G-061573-00
Stopper, wye (no. 2)
G-061574-00 or local supplier
Stopper, inspiration port (no. 3)
G-061575-00 or local supplier
Oxygen source (40 to 90 psi (275 to 620 kPa))
Local supplier
Static-dissipative field service kit (includes wrist strap,
static-dissipative mat, and ground cord)
G-061661-00 or equivalent
ESD-safe vacuum cleaner with 0.2
μ
m filter (rated for
photocopiers and laser printers)
Local supplier
Table 5-2: Test equipment required for performance verification (continued)
Description
Manufacturer and model or Covidien part
number