General information
1-10
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Table 1-2: Electromagnetic Emissions
The 700 Ventilator System is intended for use in the electromagnetic environment specified below. The customer or
the user of the 700 Ventilator System should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment–guidance
Radiated emissions
CISPR 11
Group 1
Class A
The 700 Series Ventilator System uses RF energy
only for its internal functions. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Conducted emissions
CISPR 11
Group 1
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 1-3: Electromagnetic Immunity
The 700 Series Ventilator System is intended for use in the electromagnetic environment specified below. The
customer or the user of the 700 Series Ventilator System should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic environment–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV lines/lines
± 2 kV lines/earth
± 1 kV lines/lines
± 2 kV lines/earth
Mains power quality should be
that of a typical commercial or
hospital environment.