Performance verification
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
5-11
5.
Change the settings shown in
bold
.
6.
Make the following measurements, and verify the readings are as listed below.
See
Section 5.7 15, 16, 17
.
Measure the following
Acceptable range
Tidal volume
80 to 120 mL
Respiratory rate
29 to 31 breaths/min
PEEP
0.95 to 5.05 cmH
2
O
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
0 mL
HIGH PRESSURE
90 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
20 breaths/min
TIDAL VOLUME
600 mL
PEAK FLOW
60 L/m
PLATEAU 0
s
PEEP/CPAP
15 cmH
2
O
TRIGGER SENSITIVITY
20 L/min
% O
2
21 %
Measure the following
Acceptable range
Tidal volume
530 to 670 mL
Respiratory rate
19 to 21 breaths/min
PEEP
12.56 to 17.44 cmH
2
O