Diagnostic codes
6-20
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
7145
POST
Real-time clock error
Replace controller PCB.
7146
POST
Real-time clock error
Replace controller PCB.
7147
POST
BD POST did not operate as
expected (possible
microcontroller failure).
Replace controller PCB.
7148
POST
Software error
7149
POST
Software error
7150
POST
Software error
7151
POST
Software error
7152
POST
Software error
7153
POST
General-purpose timer failure
Replace controller PCB.
7154
POST
Software error
7155
POST
BD microcontroller failure
Replace controller PCB.
7156
POST
UI POST failed.
See previous diagnostic codes in log for
corrective actions.
7157
POST
BD POST failed.
See previous diagnostic codes in log for
corrective actions.
7158
POST
BD microcontroller clock
failure
Replace controller PCB.
7159
POST
Failure to initialize UI panel
1. Replace UI display PCB.
2. Replace controller PCB.
7160
POST
Communications error (DUART
device)
Replace controller PCB.
7161
POST
Communications error (DUART
device)
Replace controller PCB.
7162
POST
Communications error (DUART
device)
Replace controller PCB.
7163
POST
Communications error (DUART
device)
Replace controller PCB.
7164
POST
Communications error (DUART
device)
Replace controller PCB.
7165
POST
Communications error (DUART
device)
Replace controller PCB.
7166
POST
Communications error (DUART
device)
Replace controller PCB.
7167
POST
Communications error (DUART
device)
Replace controller PCB.
7168
POST
General-purpose timer failure
Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action