Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-41
Figure 8-35. Exhalation assembly installed
The 700 Series
Ventilator exhalation system has been modified to reduce the chances of rain
out in the exhalation system from the humidifier generating excess humidity. This
modification consists of a change to product software (revision M or later) and the addition
of insulating foam and two isolating washers. All ventilators manufactured after April, 2002,
have this modification and the software installed. For kit part numbers, see Table 9-3.
Figure 8-36. Exhalation heater and thermistor assemblies
7-00199
Exhalation
valve
Thermistors
Heater and
clip
Remove these
screws to remove
check valve
(6 PL total)
Flow sensor
heater and
clip
Flow sensor
assembly
Cross tube
7-00200
Thermistor at
top of flow
sensor assembly
Check valve
retaining
screws
Heater and clip