Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-21
7169
POST
Ventilator did not switch to AC
after POST.
1. Verify AC is acceptable.
2. Check LEDs on BBU PCB and power
supply to determine presence of
voltages. Replace BBU PCB, power
supply, or power input component, as
required.
7170
POST
Atmospheric pressure out of
range
1. Make sure ventilator is operating
within stated environmental
specifications.
2. Replace pressure solenoid PCB.
7171
POST
Atmospheric pressure out of
range
1. Make sure ventilator is operating
within stated environmental
specifications.
2. Replace pressure solenoid PCB.
7172
POST
Analog signal on UI section of
controller PCB out of range
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
7173
POST
Analog signal on UI section of
controller PCB out of range
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
7174
POST
Analog signal on BD section of
controller PCB out of range
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
7175
POST
Cylinder pressure too low
1. Make sure inspiratory filter is not
blocked.
2. Replace pressure solenoid PCB.
7176
POST
Cylinder pressure too high
1. Make sure inspiratory filter is not
blocked.
2. Replace pressure solenoid PCB.
7177
POST
Exhalation limb temperature
too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and exhaled flow attained
a steady-state temperature.
2. Replace exhalation heater assembly.
3. Replace exhalation thermistor
assembly.
4. Replace pressure solenoid PCB.
7178
POST
Exhalation limb temperature
too high
1. Make sure room temperature is not too
high.
2. Replace exhalation heater assembly.
3. Replace exhalation thermistor
assembly.
4. Replace pressure solenoid PCB.
7179
POST
Software error
7180
POST
Software error
7181
POST
Software error
7182
POST
Multiplexer failure on pressure
solenoid PCB
Replace pressure solenoid PCB.
7183
POST
Multiplexer failure on pressure
solenoid PCB
Replace pressure solenoid PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Summary of Contents for Puritan Bennett 700 Series