Performance verification
5-16
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
5.6.10 Safety valve test (Test 9)
The safety valve test verifies the safety valve opens when the exhalation valve fails.
1.
Verify ventilator settings are as follows.
1.
Block exhalation valve exhaust port with no. 3 stopper.
2.
Block patient wye with no. 2 stopper.
3.
Verify safety valve opens when an inspiration is delivered.
NOTE:
For ventilators with software revision E or later, the safety valve opens then closes
again when it enters occlusion cycling mode. (On ventilators with earlier software than
revision E, the safety valve opens and remains open until you press ALARM RESET.)
4.
Remove stoppers.
5.
Remove test equipment, and return ventilator to original configuration.
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
5 mL
HIGH PRESSURE
15 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
6 breaths/min
TIDAL VOLUME
1500 mL
PEAK FLOW
15 L/min
PLATEAU
0 s
PEEP/CPAP
0 cmH
2
O
TRIGGER SENSITIVITY
20 L/min
% O
2
21