Diagnostic codes
6-36
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
10007
Interrupt routines
1. Software error
2. Hardware failure
1. See Section 6.2.
2. Replace controller PCB.
10008
Interrupt routines
1. Software error
2. Hardware failure
1. See Section 6.2.
2. Replace controller PCB.
10009
Interrupt routines
1. Software error
2. Hardware failure
1. See Section 6.2.
2. Replace controller PCB.
10010
Interrupt routines
Power was lost immediately
after ventilator power was
switched on.
1. Replace power supply.
2. Replace BBU PCB.
10011
Interrupt routines
Power was lost immediately
after ventilator power was
switched on.
1. Replace power supply.
2. Replace BBU PCB.
11000
SST/EST
Inspiration pressure out of
bounds
Replace pressure solenoid PCB.
11001
SST/EST
Exhalation pressure out of
bounds
1. Verify the ports were blocked correctly
during testing.
2. Replace pressure solenoid PCB.
11002
SST/EST
Exhaled flow out of bounds
1. Make sure room temperature is within
specifications. Make sure that
ventilator has warmed up and exhaled
flow attained a steady-state
temperature.
2. Replace flow sensor.
3. Replace pressure solenoid PCB.
11003-
11015
SST/EST
Software error
11016
SST/EST
Inspiration pressure out of
bounds
1. Verify the ports were blocked correctly
during testing.
2. Replace pressure solenoid PCB.
11017
SST/EST
Exhalation pressure out of
bounds
1. Verify the ports were blocked correctly
during testing.
2. Replace pressure solenoid PCB.
11018
SST/EST
Exhaled flow out of bounds
1. Make sure room temperature is within
specifications. Make sure that
ventilator has warmed up and exhaled
flow attained a steady-state
temperature.
2. Replace flow sensor.
3. Replace pressure solenoid PCB.
11019-
11131
SST/EST
Software error
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action