General information
1-12
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Table 1-4: Electromagnetic Immunity-conducted and radiated RF
The 700 Series Ventilator System is intended for use in the electromagnetic environment specified below. The
customer or the user of the The 700 Series Ventilator System should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic environment–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
outside ISM bands
a
10 Vrms
inside ISM bands
a
10 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
outside ISM bands
10 Vrms
inside ISM bands
10 V/m
80 MHz to 2.5 GHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of the 700 Series Ventilator
System, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d
0.35
P
=
d
1.2
P
=
d
1.2
P
=
d
2.3
P
=