Diagnostic codes
6-32
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
9133
Ongoing hardware
checks
Flow sensor pressure
transducer reading too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and T
xdcr
attained a steady
state.
2. Verify tubes between flow sensor and
pressure transducer are securely
connected.
3. Make sure EXHAUST port is not
blocked.
4. Replace pressure solenoid PCB.
9134
Ongoing hardware
checks
Flow sensor pressure
transducer reading too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and T
xdcr
attained a steady
state.
2. Verify tubes between flow sensor and
pressure transducer are securely
connected.
3. Make sure EXHAUST port is not
blocked.
4. Replace pressure solenoid PCB.
9135
Ongoing hardware
checks
Flow sensor pressure
transducer reading too high
1. Make sure room temperature is not too
high.
2. Verify tubes between flow sensor and
pressure transducer are securely
connected.
3. Make sure EXHAUST port is not
blocked.
4. Replace pressure solenoid PCB.
9138
Ongoing hardware
checks
+24V to UI display PCB too low
1. Replace UI display PCB.
2. Replace controller PCB.
9139
Ongoing hardware
checks
+24V to UI display PCB too
high
1. Replace UI display PCB.
2. Replace controller PCB.
9140
Ongoing hardware
checks
+5V to UI display PCB too low
1. Replace UI display PCB.
2. Replace controller PCB.
9141
Ongoing hardware
checks
+5V to UI display PCB too high
1. Replace UI display PCB.
2. Replace controller PCB.
9142
Ongoing hardware
checks
Software/UI display PCB
revision mismatch
1. Make sure UI display PCB is for a
700 Series Ventilator System.
2. Replace UI display PCB.
3. Replace controller PCB.
9143
Ongoing hardware
checks
Software/UI display PCB
revision mismatch
1. Make sure UI display PCB is for a
700 Series Ventilator System.
2. Replace UI display PCB.
3. Replace controller PCB.
9144
Ongoing hardware
checks
+1.2 V reference from pressure
solenoid PCB too low
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9145
Ongoing hardware
checks
+1.2 V reference from pressure
solenoid PCB too high
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action