S
ECTION
Self-tests
This chapter describes the 700 Series Ventilator System’s built-in self-tests,
including how to run them. It also describes how the ventilator responds when
the self-tests detect undesirable conditions. For a listing of ventilator diagnostic
codes, refer to Chapter 6. For a listing of alarm messages, refer to Chapter 7.
3.1 Introduction
The 700 Series Ventilator System has the self-test capabilities described in
Table 3-1.
Table 3-1: Self-tests
Name
Purpose
When it is run
Power-on self-
test (POST)
Verifies the integrity of the
microcontrollers, power supply,
battery, transducers, and motor.
Automatically
•
when power is applied to the
ventilator (including after a power
interruption)
•
at the start and end of SST and EST
•
when exiting from standby mode
or calibration
•
after pressure units are changed
(utility menu function)
•
after the ventilator is reset for any
reason
Short self-test
(SST)
An abbreviated version of EST to be
run primarily by the operator. Makes
detailed checks of the pneumatics and
electronics. It also characterizes system
leaks and system/tubing compliance to
compensate during breath delivery.
POST is run as part of SST.
Before patient is connected to the
ventilator or after ventilator breathing
circuit or humidifier is changed
Summary of Contents for Puritan Bennett 700 Series
Page 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Page 439: ......