Aesculap
®
ELAN 4 air motor hose 3 m with hand control GA706R or ELAN 4 air motor
hose 5 m with hand control GA707R, ELAN 4 air wall adapters GA710R to
GA715R
2.3
Indications
Note
The type and area of application depend on the applied parts and tools
selected.
ELAN 4 air motor hose with hand control
The ELAN 4 air motor hose with hand control is used in neurosurgery, ENT
surgery, OMF surgery, orthopedics and trauma surgery. It is used to con-
nect the hand-held ELAN 4 air applied parts to the compressed air source
and changes their speed on operating the hand lever.
ELAN 4 air motor system
2.4
Contraindications
The ELAN 4 air motor system is not authorized for use in the central ner-
vous system or central circulatory system.
Note
The clinical success of the use of the ELAN 4 air motor system is dependent
on the knowledge and experience of the surgeon. The surgeon must decide
which structures it is sensible to treat and take into account the safety and
warning information contained in these instructions for use.
3.
Safe handling
CAUTION
Federal law restricts this device to sale by, or on oder of a physician!
■
General risk factors associated with surgical procedures are not
described in this documentation.
■
It is the operating surgeon's responsibility to ensure that the surgical
procedure is performed correctly.
■
The operating surgeon must have a thorough understanding of both
the hands-on and conceptual aspects of the established operating
techniques.
►
Clean the new product after removing its transport packaging.
►
Prior to use, check that the product is in good working order.
►
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability:
– Use the product only according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Keep the instructions for use accessible for the user.
►
Always adhere to applicable standards.
►
Use the product within the specified pressure range only.
►
Carry out sterile processing on any reusable applied parts, motor hoses
and tools prior to use.
►
When handling, observe instructions for use TA009721 that are rele-
vant to Aesculap holder systems, see Aesculap Extranet at
https://extranet.bbraun.com
Modes of application
Separating, removing and shaping hard tis-
sue, cartilage and similar, as well as for drill-
ing, reaming and sawing bone and bone
replacement materials.
Surgical discipline/areas
of application
Neurosurgery, ENT surgery, OMF surgery,
orthopedics and trauma surgery.
DANGER
Danger of fire and explosion when operating
devices that run on oxygen!
►
Never operate the pneumatic motor with oxy-
gen!
►
Only use compressed air or nitrogen to power
pneumatic motors.
WARNING
Risk of injury and material damage due to inappro-
priate use of the product!
►
Use the product only in accordance with the
intended purpose.
WARNING
Risk of injury and damage to property due to
improper handling of the product!
This product is part of the ELAN 4 air motor sys-
tem.
►
Follow the instructions for use of all products
used.