8
Aesculap
®
ELAN 4 air motor hose 3 m with hand control GA706R or ELAN 4 air motor
hose 5 m with hand control GA707R, ELAN 4 air wall adapters GA710R to
GA715R
Disconnecting the motor hose with hand control from the wall
adapter
►
Turn the connection for wall adapter
1
of the motor hose into the mid-
dle position of the bayonet lock to ensure pressure equalization and
prevent any loud bangs.
►
Disconnect the motor hose fully from the wall adapter.
Disconnecting the wall adapter from the compressed-air source
►
Uncouple the connection for compressed-air source
8
of the wall
adapter from the compressed-air source.
6.3
Safe operation
Operation of the applied part is only possible if:
■
the ELAN 4 air applied part is released ("On" position),
■
a ELAN 4 air motor hose with hand control is connected.
Activating the applied part with the motor hose with hand control
►
Press the hand lever
7
on the motor hose with hand control.
The motor of the applied part runs.
►
Adjusting speed: Activate the hand lever
7
to a greater or lesser extent.
►
To select maximum speed: Activate the hand lever
7
fully.
7.
Validated reprocessing procedure
7.1
General safety notes
Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for reprocess-
ing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD, or possi-
ble variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
Mechanical reprocessing should be favored over manual cleaning as it
gives better and more reliable results.
Note
It should be noted that successful reprocessing of this medical device can
only be guaranteed following prior validation of the reprocessing method.
The operator/reprocessing technician is responsible for this.
The recommended chemistry was used for validation.
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
Note
For up-to-date information about reprocessing and material compatibility,
see also the Aesculap Extranet at https://extranet.bbraun.com
The validated steam sterilization procedure was carried out in the Aesculap
sterile container system.
7.2
General information
Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between appli-
cation and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive neutralizing agents or basic cleaners may result in a chemical
attack and/or fading and the laser marking becoming unreadable either
visually or by machine.
Residues containing chlorine or chlorides, e.g., in surgical residues, medi-
cines, saline solutions, and in the service water used for cleaning, disin-
fection, and sterilization, will cause corrosion damage (pitting, stress cor-
rosion) and result damage to metallic products. These must be removed by
rinsing thoroughly with demineralized water and then drying.
Perform additional drying, if necessary.
WARNING
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
the product.
WARNING
Risk of injury and material damage due to inadver-
tent activation of the applied part!
►
Secure applied parts that are not actually used
against inadvertent activation (“Off” position).
Summary of Contents for Aesculap ELAN 4
Page 3: ...A B C 10 11 6 12 4 ...