14
Aesculap
®
ELAN 4 air motor hose 3 m with hand control GA706R or ELAN 4 air motor
hose 5 m with hand control GA707R, ELAN 4 air wall adapters GA710R to
GA715R
Insufficient power
Blunt tool
Cutting edge worn
Use a new tool.
Cutting edges coated with sub-
stances, e.g. from too little rinsing
Use a new tool.
No central alignment of the crani-
otomy reamer to the dura protec-
tor bracket
Dura protector bracket bent
Poor progress during craniotomy
Use new dura protector, follow
instructions for use
(TA014473/TA014474).
Have dura protector bracket repaired
by manufacturer.
Applied part defective
Insufficient power of applied part
Follow instructions for use of the
applied part (reprocessing, mainte-
nance).
Have applied part repaired by manu-
facturer.
Large amount of heat generated
after a short time
Follow instructions for use of the
applied part (reprocessing, mainte-
nance).
Have applied part repaired by manu-
facturer.
Air resistance in motor hose with
hand control and wall adapter too
high
Use of extension hoses
Use only the motor hose with hand
control.
Hose kinked or crushed
Lay hose only through unobstructed
areas.
No free escape of exit air at dif-
fuser
Free up the openings on the diffuser.
Air supply too low
Operating pressure too low
Adjust the operating pressure to min.
6 bar.
Pressure reducer at cylinder defec-
tive
Check the pressure reducer.
Leak in hose/hose coupling
Have hose repaired by manufacturer.
Applied part is getting hot
Excessive use
Applied part heats up
Loud running noise/vibration
Follow instructions for use of the
applied part (interval operation).
Reprocessing carried out incor-
rectly
Follow instructions for use of the
applied part (reprocessing, mainte-
nance).
Shaft of applied part bent
Have applied part repaired by manu-
facturer.
Applied part defective
Have applied part repaired by manu-
facturer.
Malfunction
Cause
Detection
Remedy