Agilent B1542A User’s Guide, Edition 4
5- 17
PLSDIV Test Definitions
PLSDIV IdVd
PLSDIV IdVd
This test definition controls SMU (source monitor unit) to force the drain voltage,
controls PGU (pulse generator) to force the gate voltage pulse, controls the
oscilloscope to monitor the gate voltage and the drain voltage, calculates the drain
current, and displays the Id-Vd characteristics.
Figure 5-2
Pulsed IV Id-Vd Measurement Result Example
Device Parameters
The following parameters are available to set the test condition.
Polarity
Polarity of source output. Nch (force the specified value) or Pch
(force the negative specified value). See “Parameters” on page
5-31.
Lg
Gate length, in m.
≥
1 nm, 4 digits or 1 nm resolution.
Wg
Gate width, in m.
≥
1 nm, 4 digits or 1 nm resolution.
Temp
Temperature, in degree. 4 digits or 0.001 degree resolution.
IdMax
Maximum drain current measurement range, in A. 0 to 80 mA,
4 digits or 1
μ
A resolution.
DeviceDelay
Delay time caused by device under test, in s.
-
1
μ
s to 1
μ
s,
4 digits or 100 ps resolution. See “Parameters” on page 5-31.
,G
9G
9J
9J6WRS
9J6WDUW9J6WHS
9J6WDUW
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...