![Agilent Technologies B1542A User Manual Download Page 43](http://html.mh-extra.com/html/agilent-technologies/b1542a/b1542a_user-manual_2868004043.webp)
Agilent B1542A User’s Guide, Edition 4
2- 23
Installation
Software Installation
Updating Software
Perform the following procedure to update the pulsed IV system software.
1. Install the latest revision of the pulsed IV system software. See “Installing
Software” on page 2-21.
2. Launch EasyEXPERT or Desktop EasyEXPERT. And click the Start
EasyEXPERT button.
3. Open your workspace or create a new workspace.
4. Display the Application Test tab screen.
5. Remove checks from categories except for the Pulsed IV in the Category list.
6. Right-click a pulsed IV test definition in the Library list and click Delete
Definition of This Test to open the confirmation dialog box. And click OK to
delete the test definition from the present workspace.
Repeat this to delete all of the pulsed IV test definitions.
7. Import the pulsed IV test definitions from the following folder.
C:\Program Files\Agilent\PLSDIV\TestDefinitions
NOTE
You do not need to perform “Installing Cable Compensation Data” on page 2-21
and “Compensation” on page 3-11.
If You Remove Pulsed IV Software
Launch the Add or Remove Programs from the Control Panel and remove the
programs in the following order.
1. Agilent Technologies PLSDIV Drain Setup
2. Agilent Technologies B1542A-Opt003
3. Agilent Technologies PLSDIV A.01.02
4. Agilent Technologies PLSDIV A.01.00 or A.01.01
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...