4- 10
Agilent B1542A User’s Guide, Edition 4
Performing Measurement
Pulsed Id-Vd Measurement
Pulsed Id-Vd Measurement
Pulsed Id-Vd measurement can be performed as follows.
Figure 4-8
PLSDIV IdVd Test Definition
1. Open the PLSDIV IdVd test definition on the EasyEXPERT application test
environment.
2. Set the measurement condition to the Device Parameters, Test Parameters, and
Extended Test Parameters. For the description of the entry fields, see “PLSDIV
IdVd” on page 5-17.
3. Connect the device under test (MOSFET) as shown in “RF Probes” on page 2-4.
4. Click the Single button at the upper right corner of the EasyEXPERT screen.
Pulsed Id-Vd measurement is started.
Figure 4-9 shows the pulsed Id-Vd measurement result example.