Agilent B1542A User’s Guide, Edition 4
2- 13
Installation
Hardware Installation
Assembling Drain Switch Unit (for B1542A-024 Retrofit)
This section is required if you retrofit the option B1542A-024. This section is not
required when you install the B1542A with the option B1542A-024. The following
procedure assembles the Drain Pulse/dc Switch Unit. See Figure 2-10.
1. Unscrew and remove the top cover of the drain switch unit.
2. Remove the mounting plate inside the unit. And screw and fix it to the bias-T.
3. Connect the coaxial cables to the bias-T correctly. And fix the bias-T to the drain
switch unit.
Use a torque wrench and an open-end wrench to fasten the SMA connectors.
4. Screw and replace the top cover of the drain switch unit.
Required accessories:
•
Drain Pulse/dc Switch Unit without bias-T, 1 ea.
•
Bias-T, 1 ea.
•
Torx wrench, furnished with the option B1542A-024. This wrench is needed to
remove or fasten the screws.
Figure 2-10
Drain Pulse/dc Switch Unit
Bias-T
Drain Bias Input
(DC)
to Drain
Current
Monitor
(RF)
to DUT via
RF switch
(RF+DC)
Bias-T
Fix the mounting plate to the Bias-T.
Connect the coaxial cables to the Bias-T.
Fix the Bias-T to the unit.
DUT
Drain Current Monitor
Drain Bias Input
SMU
(RF)
(RF+DC)
(DC)
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...