4- 8
Agilent B1542A User’s Guide, Edition 4
Performing Measurement
Pulse Waveform Measurement
Pulse Waveform Measurement
Pulse waveform measurement can be performed as follows.
Figure 4-5
PLSDIV Capt Wave Test Definition
1. Open the PLSDIV Capt Wave test definition on the EasyEXPERT application
test environment.
2. Set the measurement condition to the Device Parameters, Test Parameters, and
Extended Test Parameters. For the description of the entry fields, see “PLSDIV
Capt Wave” on page 5-4.
3. Connect the device under test (MOSFET) as shown in “RF Probes” on page 2-4.
4. Click the Single button at the upper right corner of the EasyEXPERT screen.
Pulse waveform measurement is started.
Figure 4-6 is the measurement result example when DrainMonMode=Voltage,
and Figure 4-7 is the example when DrainMonMode=Current.