Agilent B1542A User’s Guide, Edition 4
5- 33
PLSDIV Test Definitions
Parameters
MeasTime
MeasTime defines the data extraction timing (see Figure 5-9) which is the timing to
extract the pulsed IV measurement data (gate voltage, drain voltage, and drain
current) from the oscilloscope measurement result (pulse waveform). Initially, this
value is set to 0.5. Changing this value may cause a measurement error. If you
change this value, check the pulse waveform and choose the point where the pulse
voltage is flat and stable.
Figure 5-9
MeasTime
SmplNum
SmplNum defines the sampling number set to the oscilloscope. Set the value in
order to display the pulse waveform on the oscilloscope screen. A small number will
improve measurement throughput.
Sampling rate of the oscilloscope is set to the value automatically calculated from
the specified pulse width and the specified sampling number.
RIIVHW
RIIVHW
*DWHSXOVH
'UDLQSXOVH
'DWDH[WUDFWLRQWLPLQJ
PHDVBWLPH[3XOVHZLGWK
9
9
3XOVHZLGWK
93XOVHDPSOLWXGH!P99
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...