6- 24
Agilent B1542A User’s Guide, Edition 4
PLSDIV TIS Commands
PlsdivUninit
vg_base
Pulse base level, in V. Numeric value,
-
4.5 to 4.5.
polarity
Polarity of pulse. Numeric value, 0 (positive) or 1 (negative).
See Figure 6-7.
where,
vg_high
>
vg_base
and 55 mV < |
vg_high
-
vg_base
| < 4.5 V must be
satisfied.
To use the gate voltage automatic adjustment function, set |
vg_high
-
vg_base
| <
4.0 V.
Figure 6-7
Pulse Level and Polarity
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivUninit
This command resets the pulsed IV test system (all instruments) and closes GPIB
communication to each instrument.
After executing this command, any Plsdiv commands except for PlsdivInit cannot
be executed. So, make sure to execute PlsdivInit before executing any command.
Filename
PlsdivUninit.exe
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
YJBEDVH
YJBKLJK
YJBEDVH
YJBKLJK
3RODULW\
3RODULW\
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...