PlsdivMeasureId
This command performs the pulsed IV measurement and returns the measurement
result. This performs drain voltage spot measurement and calculates current value.
Filename
PlsdivMeasureId.exe
Write Type
String
Read Type
List
Write String
This command needs the following input parameters in this sequence.
ranging
Measurement ranging mode. Numeric value, 0 (auto ranging) or
1 (fixed ranging).
id_range
Drain current range, in A. Numeric value, 0 to 0.08.
Ranging Mode
Auto ranging uses the best measurement range for the monitor value automatically.
The maximum range is the best range for the specified
id_range
value.
Fixed ranging uses the best range for the specified
id_range
value. This mode will
improve measurement throughput.
Read List
This command returns the following output parameters in this sequence. To read the
values, specify a vector variable in the Values field of the ReadList area. The values
are stored in the first element to the fourth element of the vector variable.
id
Drain current, calculated from
vd_actual
. Numeric value, in A.
vd_set
Drain voltage, SMU output value. Numeric value, in V.
vd_actual
Drain voltage, oscilloscope monitor value. Numeric value, in V.
conv_stat
Convergence result of the drain voltage automatic adjustment.
Numeric value, 0 (convergence condition is satisfied) or
-
1 (not
satisfied). If the automatic adjustment is off, 0 is returned. The
convergence condition is a system parameter, not
user-definable.
Auto Adjustment
This function is set by the PlsdivSetAdjustStat command, and is used to adjust the
drain voltage automatically so that the
vd_actual
value is close to the
vd
value
specified by the PlsdivForceVd command.
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...