Agilent B1542A User’s Guide, Edition 4
6- 19
PLSDIV TIS Commands
PlsdivSetDeviceDelay
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivSetDeviceDelay
This command sets the device delay time between the gate pulse and the drain pulse.
Before performing the pulsed IV measurement, figure out the device delay time and
set it by using this command. To figure out the device delay time, set up the test
system properly, connect the device under test, apply voltage pulse to the gate
terminal, and monitor the gate pulse and the drain pulse as shown in Figure 6-4.
Filename
PlsdivSetDeviceDelay.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameter.
device_delay
Delay time caused by test device, in second. Numeric value,
-
0.000001 to 0.000001.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Figure 6-4
Adjustment for Device Delay
*DWHPRQLWRUSXOVH
'UDLQPRQLWRUSXOVH
*DWHPRQLWRUSXOVH
'UDLQPRQLWRUSXOVH
'HYLFHGHOD\
%HIRUHDGMXVWPHQW
$IWHUDGMXVWPHQW
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...