PLSDIV TIS Commands
PlsdivSetSmoothing
Write String
This command needs the following input parameters in this sequence.
sampling_rate
Sampling rate. Numeric value, 0 (auto), 1 to 20 G [times/sec].
It is difficult to find the optimum value. Set
sampling_rate
=0
for an easy setup.
sampling_num
Total number of sampling. Numeric value, 0 (auto), 16 to
262144.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivSetSmoothing
This command enables or disables smoothing of the specified oscilloscope channel.
Execution
Conditions
PlsdivSetSampling command must be executed before this command.
Filename
PlsdivSetSmoothing.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameters in this sequence.
ch
Oscilloscope channel. Numeric value, 0 (channel for gate) or
1 (channel for drain).
smooth_stat
Smoothing function on or off. Numeric value, 0 (off) or 1 (on).
smooth_num
Number of measurement points used for smoothing. Numeric
value, 3 to 4001. The value must be less than the
sampling_num
parameter value set to the PlsdivSetSampling command.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Remarks
The smoothing function averages the time-series data to reduce variation caused by
high-frequency component.
Set appropriate
smooth_num
value while monitoring the waveform. Too much
smooth_num
value will lose the waveform edges, which disturbs accurate results.
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...