![Agilent Technologies B1542A User Manual Download Page 129](http://html.mh-extra.com/html/agilent-technologies/b1542a/b1542a_user-manual_2868004129.webp)
Agilent B1542A User’s Guide, Edition 4
6- 15
PLSDIV TIS Commands
PlsdivInit
Auto Adjustment
This function is set by the PlsdivSetAdjustStat command, and is used to adjust the
gate voltage automatically so that the
vg_mon
value is close to the
vg_high
value
specified by the PlsdivSetVgLevel command.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivInit
This command initializes the pulsed IV test system. This command must be
executed before any other Plsdiv command.
Filename
PlsdivInit.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameters in this sequence.
connect_check
Connection check status. Numeric value, 0 (does not perform)
or 1 (performs connection check in the PlsdivInit process).
gate_zero
Gate zero status. Numeric value, 0 (does not perform) or
1 (performs gate monitor channel offset measurement in the
PlsdivInit process).
Gate Zero
This function is used to measure the offset data of the gate voltage monitor path and
update the offset value in a system file. The offset value is used until the next
execution. Before performing the offset measurement, set up the test system
measurement path properly and open the measurement terminals.
Connection Check
This function is used to confirm if the divider, terminator, and bias-T are connected
properly. For the proper connections, see Figure 6-2 and Figure 6-3. The connection
check can detect the following improper connections.
•
divider is not connected
•
terminator is not connected
•
T-adapter is connected instead of divider
•
bias-T’s DC+RF connector is connected to the gate monitor channel
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...