Agilent B1542A User’s Guide, Edition 4
2- 15
Installation
Hardware Installation
Making Measurement Path for Gate (for B1542A-024)
Connect the following accessories as shown in Figure 2-12. And fix the gate
pulse/dc switch unit to the appropriate place. Use a torque wrench and an open-end
wrench to fasten the SMA connectors.
Required accessories:
•
BNC(m)-SMA(f) adapter, 1 ea., for pulse generator
•
BNC(m)-SMA(f) adapter, precision type, 1 ea., for oscilloscope
•
BNC cable, 1 ea.
•
SMA cable, 0.3 m, 1 ea.
•
SMA cable, 1.5 m, 1 ea.
•
Triaxial cable, 1 ea.
Figure 2-12
Measurement Path for Gate
Scope
DC source monitor
Triaxial cable
Pulse generator
to SMU Force
for gate voltage
at DC test
Gate pulse/dc switch unit
SMA cable, 0.3 m
BNC cable
SMA cable, 1.5 m
to input channel
for gate (*2)
Gate Pulse Monitor
SMU
to output channel for gate pulse (*1)
(*1) Use BNC(m)-SMA(f) adapter
for pulse generator.
(*2) Use BNC(m)-SMA(f) adapter
precision type for oscilloscope.
to TRIGGER OUT
to AUX Trig In
Gate Pulse Input
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...