Agilent B1542A User’s Guide, Edition 4
6- 11
PLSDIV TIS Commands
PlsdivCaptureMonData
PlsdivCaptureMonData
This command reads the oscilloscope vertical data (monitor data). It returns the gate
voltage data, drain voltage data, or drain current data selected by the
term
and
drain_mode
parameters.
Execution
Conditions
PlsdivSetSampling and PlsdivMeasureId must be executed before this command.
The PlsdivSetSampling command
sampling_num
parameter value must not be 0.
Filename
PlsdivCaptureMonData.exe
Write Type
String
Read Type
List
Write String
This command needs the following input parameters in this sequence.
term
Terminal. Numeric value, 0 (gate) or 1 (drain).
sampling_num
Sampling number of oscilloscope. Numeric value, 16 to
262144. The value must be the same as the
sampling_num
parameter value set to the PlsdivSetSampling command.
drain_mode
Drain data type. Numeric value, 0 (voltage) or 1 (current).
If
term
=0, this command returns the gate voltage data.
If
term
=1 and
drain_mode
=0, this command returns the drain voltage data.
If
term
=1 and
drain_mode
=1, this command returns the drain current data.
Read List
This command returns the following output parameter.
read_data
Oscilloscope vertical data (monitor data). Vector data. The
number of returned data is
sampling_num
if the smoothing is
off or the number of data displayed on the oscilloscope screen if
the smoothing is on. The smoothing on/off is set by the
PlsdivSetSmoothing command.
Response
Number of returned data. Numeric value.
Summary of Contents for B1542A
Page 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Page 21: ...2 Installation ...
Page 49: ...3 Performing System Setup and Compensation ...
Page 65: ...4 Performing Measurement ...
Page 79: ...5 PLSDIV Test Definitions ...
Page 115: ...6 PLSDIV TIS Commands ...
Page 139: ...7 Status Code and Error Messages ...