
220
Purged/Packed Inlet
Replacement procedures
8 of 20
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Replacing an EPC flow manifold
WARNING
Before proceeding, turn off the main power switch and let the heated zones
cool. Follow ESD precautions.
All EPC inlets and the ECD detector in the 6890 GC use Type 1 flow modules.
1.
Shut off the main gas supply to the manifold and remove the 1/8-inch
Swagelok fitting for the gas supply.
2.
Remove the plastic detector cover and the plastic pneumatics cover.
3.
Remove the metal RFI shield, and the rear cover on the back of the GC.
4.
Disconnect the ribbon cable for the manifold from the main EPC board.
You may have to remove the adjacent ribbon cable also.
Figure 220-8
Removing the EPC flow manifold
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......