1120
Valve Box
Installing the valve box
6 of 10
Jun 2001
Valves
Agilent 6890 Gas Chromatograph Service Manual
Figure 1120-7
Assembling the valve box top
12. While wearing a pair of protective gloves and holding the valve box top
over a waste receptacle, punch or cut out the insulation from the
appropriate holes.
13. Use two Torx T-20 screws to secure the valve box top over the installed
valves, making sure the heater/sensor leads are routed under the
appropriate cutout(s).
Make sure the valve box top is oriented correctly with holes punched out
over the installed valves.
Retainer plate
Insulation
Valve box top