7 of 8
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Wiring Harnesses
1240
Temperature sensor resistance
Temperature sensor resistance
The approximate resistance of a temperature sensor is:
R = 100 + (.35
×
t)
where R is resistance in ohms and t is sensor temperature in °C.
Table 1240-1
Temperature Sensor Resistance by Heater Temperature
Figure 1240-3
Chart of temperature sensor resistances by heater temperature
°C
+0°
+10°
+20°
+30°
+40°
+50°
+60°
+70°
+80°
+90°
+100°
0°
100.00
103.90
107.79
111.67
115.54
119.40
123.24
127.07
130.89
134.70
138.50
100°
138.50
142.28
146.06
149.82
153.57
157.32
161.04
164.76
168.47
172.16
175.84
200°
175.84
179.51
183.17
186.82
190.46
194.08
197.70
201.30
204.88
208.46
212.03
300°
212.03
215.58
219.13
222.66
226.18
229.69
233.19
236.67
240.15
243.61
247.06
400°
247.06
250.50
253.34
257.34
260.75
264.14
267.52
270.89
274.25
277.60
280.93
281W
250W
200W
150W
100W
0
100
°
200
°
300
°
400
°
500
°
TEMPERATURE (
°
C
)
R
S
I
S
T
A
N
C
E
E
(W)
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......