341
Micro-cell Electron Capture Detector (
µ
-ECD)
Maintaining a
µ
-ECD detector
26 of 26
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
It is good practice to monitor the progress of the thermal cleaning by
plotting the
µ
-ECD signal. Over time, the signal baseline signal should
change as shown in Figure 341-18.
Figure 341-18
Time
7.
Check the ECD “output” value from the GC display. It should be lower
than the first reading.
Performing a radioactivity leak test (wipe test)
Micro-cell ECDs must be tested for radioactive leakage at least every 6 months.
Records of tests and results must be maintained for possible inspection by
the Nuclear Regulatory Commission and/or responsible state agency. More
frequent tests may be conducted when necessary.
The procedure used is the wipe test. A Wipe Test Kit (part no. 18713-60050)
is supplied with each new
µ
-ECD. Refer to the information card supplied in
the Wipe Test Kit for instructions on performing the wipe test.
Start of thermal cleaning
Baseline before cleaning
Baseline after
cleaning
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......