21 of 26
Jun 2001
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms
610
Microcell ECD checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
6890 Analog output
Noise ([N1])
Measured at [6890 Range 0],
[3396 Attn. 3, Chart speed 1, zero 50]
<37 mm for 1 min. measurement
Signal
6890 Range 6
3396
Zero
10
Pk wd
0.04
Att 2^
5
Thresh
5
Cht sp
1
Stop time
11.0
Ar Rej
100000
Area, Lindane peak
47,950 × [N1]
Agilent ChemStation
Noise ([N3])
Measured by ChemStation, ASTM [N3]
< 3 Hz
Signal
Area, Lindane peak
7500 × [N3]