1120
Valve Box
Installing the valve box
4 of 10
Jun 2001
Valves
Agilent 6890 Gas Chromatograph Service Manual
7.
Insert the valve(s) into the appropriate hole(s) in the heater block and
secure each valve with two long Torx T-10 screws.
Figure 1120-5
Installing valves
8.
Plumb the valves in the appropriate configuration. A variety of valve
configurations are diagrammed in the
Typical Valve Configurations
section in this chapter.
9.
Orient all of the valves in the OFF (CCW) position.