210
Split/Splitless Inlet
Leak testing—EPC and manual inlets
20 of 30
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 210-14
Capping the bottom of the inlet and septum purge vent
Note
Make sure that the carrier gas source pressure is at least 35 psi. Carrier source
pressure should always be at least 10 psi greater than the desired inlet
pressure.
Manual
EPC
Split vent
Septum purge vent
Septum purge vent