Agilent 6890 Gas Chromatograph Service Manual
510
Environmental
Temperature and humidity ranges
Operating the GC within the recommended ranges insures optimum
instrument performance and lifetime.
After exposing the GC to extremes of temperature or humidity, allow 15
minutes for it to return to the recommended ranges.
Ventilation requirements
The GC is cooled by convection: air enters vents in the side panels and
underneath the instrument. Warmed air exits through slots in the top, rear,
and side panels. Do not obstruct air flow around the instrument.
Caution
For proper cooling and general safety, always operate the instrument with
cover panels properly installed.
Venting oven exhaust
Hot air (up to 450°C) from the oven exits through a vent in the rear. Allow at
least 20 cm (10 inches) clearance behind the instrument to dissipate this air.
Recommended temperature range
Temperature range
20 to 27
°
C
5 to 40
°
C
Recommended humidity range
Humidity range
50 to 60%
Up to 31
°
C, 5 to 80%
At 40
°
C, 5 to 50%
Recommended altitude range
Up to 2000 m
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......