350
Flame Photometric Detector (FPD)
Replacement procedures
10 of 36
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
6.
Remove the four screws in the PMT adapter flange and remove the flange.
Use care as a quartz window is exposed and may fall out. Clean the
window using lens tissue.
Figure 350-5
FPD disassembly
O-ring,
Viton
0-ring
Window, second heat shield
0-ring
Flange adapter
Flange ring
Clamp