430
Power/Electronics Replacement
Replacing the main board
4 of 26
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
7.
If an MIO card is installed, remove the card and the support bracket as
described below:
a.
Loosen the two spring-loaded captivated screws on the back of the
instrument and slide the card out.
b.
Remove the MIO support bracket.
c.
Remove the three screws on the back of the instrument and the one
screw in the standoff on the side of the instrument.
Figure 430-3
Removing the card and three screws on the back of the GC