420
Oven and Temperature Control
Replacing the oven shroud assembly
12 of 20
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 420-6
Removing the shroud
6.
From inside the oven, draw the heater and sensor leads into the oven and
remove the fan shroud.
7.
Reassembly is the reverse of removal. Note that when reconnecting the
heater leads to P3 and P4 on the AC board, either of the leads may be
connected to either of the connectors.
Loosen
Remove