410
Covers and Fans
Removing the pneumatics top cover
8 of 20
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Removing the pneumatics top cover
The large plastic cover over the pneumatics area on the top, rear of the GC
simply lifts off. To replace the cover, line up the holes in the top of the cover
with any pneumatics fittings and regulator dials and snap it back into place.
Removing the pneumatics RFI shielding
An aluminum cover shields the top of the pneumatics area at the top, rear of
the instrument.
1.
Remove the plastic pneumatics cover and remove the one Torx T-20 screw
from the top of the metal RFI shielding.
2.
From the rear of the GC, slide the RFI shielding to the left, out from under
the locking tabs, and lift it off.
Figure 410-8
Removing the RFI shielding