
610
Checkout Chromatograms
Microcell ECD checkout conditions and chromatogram
18 of 26
Jun 2001
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Microcell ECD checkout conditions and chromatogram
Column and sample
Type
HP-5 30m × 0.32 mm × 0.25 µm PN 19091J-413
Sample
ECD Checkout PN 18713-60040
Injection volume
1 µL
Inlet
Temperature
200°C Purged/Packed
250°C
Split/Splitless
Oven TrackCool On-Column
80°C
PTV (see below)
Inlet pressure
25 psi (Constant pressure for EPC inlets,
helium)
Split/Splitless Only
Mode
Splitless
Purge flow
60 mL/min
Purge time
0.75 min
PTV—Splitless Mode
Inlet temperature
80°C
Initial time
0.1 min
Rate 1
720°C/min
Final temp 1
350°C
Final time 1
2 min
Rate 2
100°C/min
Final temp 2
250°C
Final time 2
0 min
Inlet pressure
25 psi (const. press. for EPC inlets)
Purge time
0.75 min
Purge flow
60 mL/min
Detector
Temperature
300
°
C
Const. Makeup Flow, N
2
30 mL/min
Offset
Should be <400 display units
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......