Agilent 6890 Gas Chromatograph Service Manual
Power/Electronics Replacement
430
Replacing the battery
Replacing the battery
WARNING
Before proceeding, turn off the main power switch.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
Removing the battery while the instrument is unplugged erases all user
entered setpoints, methods and sequences.
The 3V battery for the 6890 GC supplies the power to retain user entries in
memory, such as programmed methods and sequences, in the event of a power
failure or when the instrument is unplugged. The battery is located in the
lower left corner of the main board.
To replace the battery, slide the battery down and pull the battery out of the
bottom first.
Figure 430-10
Location of the battery on the main board
Hole for
Oven Access
Cutout
Beeper
P11
P1
P2
P12
P13
P17
GND
P15
DSP
Gate
Array
CPU
GND
4 ROM
Sockets
3V
Lithium
battery
Gate
Array
P16
P21
P22
P3
J1
J2
JP1
JP2
J4
J6
J5
F2 F1
F4 F3
P18
+24V
–24V
+15V
–15V
+5v
Capacitor
Capacitors
J8
transformer
cable
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......