210
Split/Splitless Inlet
Replacement procedures
8 of 30
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Removing the inlet
WARNING
Turn off the oven and turn off the heater of the inlet you are replacing and
let them cool down. Turn off all detector flows. Turn off the carrier gas supply
pressure, then turn off the main power switch and unplug the power cord.
1.
From inside the oven, remove the column and column fitting from the
bottom of the inlet.
2.
Remove the left side cover from the GC.
3.
From the top of the 6890 GC, remove the blue inlet carrier cover (or the
tray bracket, if installed).
4.
Unclip the heater/sensor leads from the connector to the left of the inlet
carrier.
5.
Unscrew the top insert assembly (large inlet nut and flow lines on the
top of the inlet) that has the carrier gas and septum purge lines plumbed
to it.
6.
Raise this assembly up and away from the inlet
Figure 210-5
Removing the inlet
Purge
Carrier in
Top insert assembly
Split vent flow line nut
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......