Agilent 6890 Gas Chromatograph Service Manual
1230 Power Supply
Setting the instrument power configuration
There are seven possible line voltage power configurations for the 6890 GC.
Table 1230-1
Voltage Configuration Information
To change the power configuration for the instrument, you must install the
appropriate types of the following components:
•
Line voltage configuration plug
•
Ceramic fuses on the AC power board
•
Oven shroud assembly
These three components are explained in the following topics.
Voltage
(–10%, +5%)
Frequency
(Hz)
Maximum power
consumption (VA)
Power line requirement
Oven type
120 V
48-66
2,250
20-amp dedicated
receptacle
Regular
200 V
48-66
2,950
15-amp dedicated
receptacle
Fast-heating
208 V
48-66
2,950
15-amp dedicated
receptacle
Fast-heating
220 V
48-66
2,950
15-amp dedicated
receptacle
Fast-heating
230 V
48-66
2,250
10-amp dedicated
receptacle
Regular
230 V
48-66
2,950
16-amp dedicated
receptacle
Fast-heating
240 V
48-66
2,950
13- or 16-amp dedicated
receptacle
Fast-heating
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......