Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes
830
Warning 136—Front detector title mismatch
Warning 136—Front detector title mismatch
Warning 137—Back detector title mismatch
The title for the detector does not match the title stored in its pneumatics
module. No action is taken. If the detector is a different type from that stored
in the pneumatics module, a hardware error will be declared.
Warning 138—Front injector/inlet mismatch
Warning 139—Back injector/inlet mismatch
The injector is not compatible with the inlet it is intended to use.
Warning 140—Front inlet Saver time changed
Warning 141—Back inlet Saver time changed
While using a Volatiles Interface inlet, the Sample End time is set less than
the inlet purge time. The purge time has been changed to the Sample End time
+ 1 to avoid the conflict.
Warning 142—Front inlet Purge time changed
Warning 143—Back inlet Purge time changed
While using a Volatiles Interface inlet, the Sample End time is set greater than
the inlet gas saver time. The gas saver time has been changed to the Sample
End time + 1 to avoid the conflict.
Warning 144—OEM Config error(s)
One or both detectors installed are not the same as the one(s) in the method.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......