Agilent 6890 Gas Chromatograph Service Manual
320
Nitrogen Phosphorus Detector
(NPD)
Theory of operation
The NPD (also called a thermionic detector) uses a jet and collector similar
in appearance to a Flame Ionization Detector. In an NPD, however, ions of
alkali metal are introduced into a flame where hydrogen and air flows are
less than those for an FID, minimizing the normal hydrocarbon ionizations,
and increasing the ionization of nitrogen or phosphorous compounds. This
causes the NPD to be both sensitive and selective for organic compounds
containing nitrogen
and/or phosphorous. This thermionic source efficiently ionizes nitrogen and
phosphorous containing organic molecules. Ions are collected and the
resulting current measured for the chromatogram.
Figure 320-1
The Nitrogen Phosphorus Detector (NPD)
EPC detector
The EPC NPD has three pneumatic supply lines for air, hydrogen and makeup
gas. Each line has a filter frit, a proportional valve controlled by a pressure
Bead drive cable
Eletrometer
Electrometer ribbon cable
Pneumatic lines
Detector body
Heater/sensor cable
(under cover)
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......